When and how to fill NOTCam with LN2
It is the task of the person on duty to check the NOTCam temperature
status and possibly fill NOTCam with LN2.
- NOTCam should be filled when you get the email alert which is
sent out to duty staff if/when the detector reaches a temperature
above -193.0 degrees.
- Instructions for
How to fill NOTCam with LN2
More details about NOTCam filling
NOTCam should be filled every morning whenever it is mounted
on the telescope to make sure it reaches operating temperature
for the coming night. If you are observing, fill it before starting
the notcam-calibs in the morning, and if you are supporting, fill
it when you arrive at the telescope the next morning.
During storage NOTCam is kept cold by the PTR and regular LN2
fillings. Whenever needed, email alerts will be sent to staff
that the detector temperature has reached the limit when
it is time to fill LN2. This limit has been set very
conservatively. If the PTR works well, you can typically wait until
the next morning if the first email arrives in the early night.
NB! If the PTR is not working, you should act quickly!
You can check the NOTCam temperatures via the
TMS monitoring on the web. Sensors 500-505.
You can also do ssh -l obs selena from any terminal using the usual
password. Check the temperature and pressure log file with the command
tail -f /home/software/notcam/NOTCAM_temp. Each minute four
different temperatures and the pressure is updated. The first
temperature is the cold table, the next is the outer vessel (typically
reflecting the dome temperature), the third is the center wheel section,
and the forth is the array temperature. The last value is the pressure
in mbar.
Mon Apr 7 10:17:47 2003 1049707067 -201.60 3.80 -197.30 -196.30 4.16e-05
|