- Telescope start-up:
Make sure that the telescope power is on, hatch(es) and mirror cover
are open (see cookbook)
- Make sure that the STANCAM
sequencer has been started.
- Go to the script directory (e.g. cd scripts/45-678), if needed.
- Note the Cassegrain instrument setup, especially the focus
value, if you will use the Cassegrain instrument after StanCam
- Telescope setup:
run script tcs.setup-tel-stancam on a StanCam sequencer window.
This will:
tcs.ag-off
tcs.ccd-filter 0 (STANCAM no filter)
tcs.camera-probe-ccd
tcs.tcs-command "instrument-name STANCAM"
tcs.tcs-command "field-rotation 0"
tcs.offset-size 10 10
tcs.probe-step-size 5 5
tcs.tv-focus
tcs.tcs-command "box-at-center"
tcs.background-box-position 400 400
tcs.focus-position
Note, all above commands can be given from the TCS as well, e.g.
focus-position
- Focus the telescope, if needed
- Preset the telescope to the target
- Choose the desired
filter
e.g. ccd-filter 4 when using the TCS
or tcs.ccd-filter 4 when using the
sequencer
- Start autoguiding, if needed, (e.g. ag-on,
see the TCS pages)
- Adjust the telescope focus, if needed (e.g.
tcs.focus-position 23080 or tcs.focus-delta 40
- Start integrating (e.g. expose 120) or run the script,
(see the
StanCam sequencer commands )