Somewhat similar to FIES, we also had problems with maintaining the StanCam detector cold. After some investigation it was concluded that this was the common feature of the CryoTiger cooler for unknown reasons loosing cooling capacity. For StanCam it is only possible to see if the CryoTiger is functioning properly by looking at the detector temperature which is 0.1 degree below its set temperature of -95.0 degrees Celsius.
One of the potential causes for the CryoTiger loosing cooling capacity
is the dewar loosing vacuum. Ever since StanCam was mounted we have
been unable to obtain its pressure without having to go in to the dome
and read it off a digital display. It was found that the pressure
sensor display is actually a complete controller unit, and has an
RS-232 port through which it is possible to obtain the pressure
reading. Via a system using an Ethernet/RS-232 (Moxa box) interface it
is now possible to see the pressure in the data base. The main
limitation is that the sensor only goes down to millibar which
implies that under normal circumstances it will only provide an upper
limit, but at least we have now a way of automatically detecting any
major loss of vacuum such that any corrective action can be taken in a
timely manner.
Thomas Augusteijn 2012-02-21